Publication:
Ultra-Compact Camera-Integrated Deep-Implanted Two-Port MIMO Antenna for High-Data-Rate Wireless Capsule Endoscopy
| cris.virtual.department | #PLACEHOLDER_PARENT_METADATA_VALUE# | |
| cris.virtual.orcid | 0000-0003-0414-9147 | |
| cris.virtualsource.department | 1697b691-7f8b-4d10-bef3-42fa20421383 | |
| cris.virtualsource.orcid | 1697b691-7f8b-4d10-bef3-42fa20421383 | |
| dc.contributor.author | Qamar, Muhammad | |
| dc.contributor.author | Kapusuz, Kamil Yavuz | |
| dc.contributor.author | Carslake, Lawrence | |
| dc.contributor.author | Loh, Tian-Hong | |
| dc.contributor.author | Thaha, Mohamed A. | |
| dc.contributor.author | Alomainy, Akram | |
| dc.contributor.imecauthor | Kapusuz, Kamil Yavuz | |
| dc.contributor.orcidimec | Kapusuz, Kamil Yavuz::0000-0003-0414-9147 | |
| dc.date.accessioned | 2024-12-17T10:55:53Z | |
| dc.date.available | 2024-12-16T16:59:30Z | |
| dc.date.available | 2024-12-17T10:55:53Z | |
| dc.date.issued | 2025 | |
| dc.description.abstract | This study proposes a novel high-performance, two-element camera-integrated antenna array for the [401-406] MHz medical device radio communications service (MedRadio) band. The array is designed for integration inside a capsule to establish a stable, high data-rate multiple-input multiple-output (MIMO) wireless communication link for wireless endoscopy. The compact antenna (π× 1.125 ×1.49 mm3) uses stacked meandered resonators, nestled within an annular ring-shaped planar substrate, offering over 7 MHz bandwidth and 41 dB isolation. It achieves a 78 Mbps data rate over 1.6 m in muscle tissue with a specific absorption rate of 2 W/kg at 20 mW power. Measurements show an ECC below 0.001, diversity gain over 9.99 dB, and a channel capacity of 10 bps/Hz at 20 dB SNR. To the best of our knowledge, this is the first dual-camera-integrated MIMO antenna system for the [401-406] MHz MedRadio band tested in an implantable device. | |
| dc.description.wosFundingText | This work was supported in part by the 21NRM03 MEWS Project, which receives funding from the European Partnership on Metrology, in part by the European Unions Horizon Europe Research and Innovation Programme, and in part by the participating nations. | |
| dc.identifier.doi | 10.1109/JERM.2024.3501385 | |
| dc.identifier.issn | 2469-7257 | |
| dc.identifier.uri | https://imec-publications.be/handle/20.500.12860/44978 | |
| dc.publisher | IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC | |
| dc.source.beginpage | 293 | |
| dc.source.endpage | 301 | |
| dc.source.issue | 3 | |
| dc.source.journal | IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY | |
| dc.source.numberofpages | 9 | |
| dc.source.volume | 9 | |
| dc.subject.keywords | POWER TRANSFER SYSTEM | |
| dc.subject.keywords | SMALL-BOWEL | |
| dc.subject.keywords | DESIGN | |
| dc.subject.keywords | PROPAGATION | |
| dc.subject.keywords | EFFICIENCY | |
| dc.subject.keywords | BANDS | |
| dc.subject.keywords | LINK | |
| dc.subject.keywords | MHZ | |
| dc.title | Ultra-Compact Camera-Integrated Deep-Implanted Two-Port MIMO Antenna for High-Data-Rate Wireless Capsule Endoscopy | |
| dc.type | Journal article | |
| dspace.entity.type | Publication | |
| Files | Original bundle
| |
| Publication available in collections: |